VESSEL SEALING UNIT
Combo unit of ESU and Vessel Sealer
Intelligent, Smart, and Clinical
Micro-controller design
PPREM technology (Pad Patient Return Electrode Monitoring)
Surgical Assistance (99 Programs)
Error Indication
Endo Cut (Pulse Cut)
Bipolar Sealing (150w)
Four Bipolar Modes, Max. power of 150 W
Four Unipolar CUT modes, Max. power of 300 W
Four Unipolar COAG modes, Max. power of 150 W
Auto stops the RF energy upon completion of Seal
Separate Footswitch for Unipolar and Bipolar
With Bi-Clamp for Open surgery
Optional Disposable/Reusable Laparoscopic attachments
ESU Applications: | Vessel Sealer Applications: | ||
• Gynecology • Urology( TURP ) • Laproscopic Surgeries • Endoscopy ( ERCP ) • Oncosurgery • Heart & Chest surgery • Neurosurgery • Spine • Arthroscopy • All other general surgical procedures |
• Abdominal Hysterectomy • Adrenalectomy • Appendectomy • Colectomy • Colon Resection • Cystectomy • Gastrectomy • Gastric Bypass • Hemorrhoidectomy • Whipple |
• Lap-Assisted Vaginal Hysterectomy (LAVH) • Lap-Cholecystectomy • Liver Resection • Nephrectomy • Nissen Fundoplication • Radical Hysterectomy • Radical Prostatectomy • Splenectomy • Salpingo-oophorectomy • Vaginal Hysterectomy |
Technical Specifications |
|
Monopolar CUT | |
PURE CUT | 480 Khz sinusoid, 300 W @ 300 E |
BLEND1 | 480 Khz sinusoid with 70% duty cycle, 250 W @ 300 E |
BLEND2 | 480 Khz sinusoid with 55% duty cycle, 200 W @ 300 E |
ENDO CUT | 480 Khz sinusoid delivered in bursts, 300 W @ 300 E |
Monopolar COAG | |
FORCE | 480Khz damped sinusoid bursts, 150 W @ 500 E |
SPRAY | 480Khz sinusoid bursts, 120 W @ 500 E |
FULGURATE | 480Khz damped sinusoid bursts, 150 W @ 500 E |
DESICCATE | 480 Khz sinusoid with very low max. voltage, 80 W @ 500 E |
BIPOLAR | |
AUTO | 480 Khz sinusoid with lower output voltage, 80 W @ 100 E |
CUT | 480 Khz sinusoid with medium output voltage, 80 W @ 100 E |
COAG | 480 Khz sinusoid with lower output voltage, 80 W @ 100 E |
SEAL | 480 Khz sinusoid with higher output voltage, 150 W @ 100 E |
Patient Electrode monitoring System (Resistance range) | Dual area PREM patient return Electrode: 10 – 135 ohms |
Single Patient Return electrode: < 20 ohms | |
Power Deviation | less than 15% or 5 watts, whichever is greater |
Dimensions(L X W X H) | 410 x 350 x 120 mm |
Weight | < 7 kg |
Power supply | 180 – 270 Vac , 50/60 Hz |
Max. Power Input | 1000VA, Fuse : 5A |
Standard Accessories: | Optional Accessories: | |
• Monopolar handle with silicon cable with electrode set • Bipolar forcep with silicon bipolar cable • Silicon patient plate with cable • Double Pedal Footswitch • Disposable Hand switch • Mains cord • Unit Cover |
• Reusable Hand switch • Bionet Bipolar forcep • Long 9” Bipolar forcep • Small 4” Bipolar forcep • Angled tip Bipolar forcep • Monopolar forcep with cable |
GEM 400 B+
Intelligent, Smart, and Clinical
Micro-controller Design
PPREM technology (Pad Patient Return Electrode Monitoring)
Surgical Assistance (10 Programs)
Error Indication (Displays error messages on the display)
Endo Cut (Pulse Cut)
Four Bipolar Modes with Bipolar CUT, Max. power of 150 W
Four Monopolar CUT modes, Max. power of 400 W
Four Monopolar COAG modes, Max. power of 150 W
Separate Display for Monopolar CUT, COAG, and BIPOLAR
Separate Footswitch for Unipolar and Bipolar
ESU Applications: | ||
• Gynecology • Urology( TURP ) • Laproscopic Surgeries • Endoscopy ( ERCP ) |
• Oncosurgery • Heart & Chest surgery • Neurosurgery |
• Spine • Arthroscopy • All other general surgical procedures |
Technical Specifications |
|
Monopolar CUT | |
LOW | 480 Khz sinusoid with low max. voltage, 300 W @ 300 E |
PURE CUT | 480 Khz sinusoid, 400 W @ 300 E |
BLEND1 | 480 Khz sinusoid with 70% duty cycle, 250 W @ 300 E |
BLEND2 | 480 Khz sinusoid with 55% duty cycle, 200 W @ 300 E |
Monopolar COAG | |
SPRAY | 480Khz sinusoid bursts, 120 W @ 500 E |
FULGURATE1 | 480Khz damped sinusoid bursts, 150 W @ 500 E |
FULGURATE2 | 480Khz damped sinusoid bursts, 150 W @ 500 E |
DESICCATE | 480 Khz sinusoid with very low max. voltage, 80 W @ 500 E |
BIPOLAR | |
MICRO | 480 Khz sinusoid with lower output voltage, 80 W @ 100 E |
MACRO | 480 Khz sinusoid with higher output voltage, 80 W @ 100 E |
STANDARD | 480 Khz sinusoid with medium output voltage, 80 W @ 100 E |
CUT | 480 Khz sinusoid with high output voltage, 150 W @ 100 E |
Patient Electrode monitoring System (Resistance range) | Dual area PREM patient return Electrode: 10 – 135 ohms |
Single Patient Return electrode: < 20 ohms | |
Power Deviation | less than 15% or 5 watts, whichever is greater |
Dimensions(L X W X H) | 400 x 330 x 120 mm |
Weight | < 7 kg |
Power supply | 180 – 270 Vac , 50/60 Hz |
Max. Power Input | 1000VA, Fuse : 5A |
Standard Accessories: | Optional Accessories: | |
• Monopolar handle with silicon cable with electrode set • Bipolar forcep with silicon bipolar cable • Silicon patient plate with cable • Double Pedal Footswitch • Disposable Hand switch • Mains cord • Unit Cover |
• Reusable Hand switch • Bionet Bipolar forcep • Long 9” Bipolar forcep • Small 4” Bipolar forcep • Angled tip Bipolar forcep • Monopolar forcep with cable |
GEM 300
Intelligent, Smart, and Clinical
Micro-controller Design
PPREM technology (Pad Patient Return Electrode Monitoring)
Surgical Assistance (10 Programs)
Error Indication (Displays error messages on the display)
Endo Cut (Pulse Cut)
Three Bipolar Modes with Bipolar CUT, Max. power of 150 W
Three Monopolar CUT modes, Max. power of 300 W
Three Monopolar COAG modes, Max. power of 150 W
Separate Display for Monopolar CUT, COAG, and BIPOLAR
Separate Footswitch for Unipolar and Bipolar
ESU Applications: | ||
• Gynecology • Urology( TURP ) • Laproscopic Surgeries • Endoscopy ( ERCP ) |
• Oncosurgery • Heart & Chest surgery • Neurosurgery |
• Spine • Arthroscopy • All other general surgical procedures |
Technical Specifications |
|
Monopolar CUT | |
PURE CUT | 480 Khz sinusoid, 300 W @ 300 E |
BLEND1 | 480 Khz sinusoid with 70% duty cycle, 250 W @ 300 E |
BLEND2 | 480 Khz sinusoid with 55% duty cycle, 200 W @ 300 E |
Monopolar COAG | |
SPRAY | 480Khz sinusoid bursts, 120 W @ 500 E |
FULGURATE | 480Khz damped sinusoid bursts, 150 W @ 500 E |
DESICCATE | 480 Khz sinusoid with very low max. voltage, 80 W @ 500 E |
BIPOLAR | |
MICRO | 480 Khz sinusoid with lower output voltage, 80 W @ 100 E |
MACRO | 480 Khz sinusoid with higher output voltage, 80 W @ 100 E |
CUT | 480 Khz sinusoid with high output voltage, 150 W @ 100 E |
Patient Electrode monitoring System (Resistance range) | Dual area PREM patient return Electrode: 10 – 135 ohms |
Single Patient Return electrode: < 20 ohms | |
Power Deviation | less than 15% or 5 watts, whichever is greater |
Dimensions(L X W X H) | 400 x 330 x 120 mm |
Weight | < 7 kg |
Power supply | 180 – 270 Vac , 50/60 Hz |
Max. Power Input | 1000VA, Fuse : 5A |
Standard Accessories: | Optional Accessories: | |
• Monopolar handle with silicon cable with electrode set • Bipolar forcep with silicon bipolar cable • Silicon patient plate with cable • Double Pedal Footswitch • Disposable Hand switch • Mains cord • Unit Cover |
• Reusable Hand switch • Bionet Bipolar forcep • Long 9” Bipolar forcep • Small 4” Bipolar forcep • Angled tip Bipolar forcep • Monopolar forcep with cable |
GEM 200
Intelligent, Smart, and Clinical
Micro-controller Design
PPREM technology (Pad Patient Return Electrode Monitoring)
Surgical Assistance (10 Programs)
Error Indication (Displays error messages on the display)
Endo Cut (Pulse Cut)
Two Bipolar Modes with Bipolar CUT, Max. power of 150 W
Three Monopolar CUT modes, Max. power of 200 W
Three Monopolar COAG modes, Max. power of 150 W
ESU Applications: | ||
• Gynecology • Endoscopy ( ERCP ) • Oncosurgery |
• Heart & Chest surgery • Neurosurgery • Spine |
• Arthroscopy • All other general surgical procedures |
Technical Specifications |
|
Monopolar CUT | |
PURE CUT | 480 Khz sinusoid, 200 W @ 300 E |
BLEND1 | 480 Khz sinusoid with 70% duty cycle, 150 W @ 300 E |
BLEND2 | 480 Khz sinusoid with 55% duty cycle, 120 W @ 300 E |
Monopolar COAG | |
SPRAY | 480Khz sinusoid bursts, 120 W @ 500 E |
FULGURATE | 480Khz damped sinusoid bursts, 150 W @ 500 E |
DESICCATE | 480 Khz sinusoid with very low max. voltage, 80 W @ 500 E |
BIPOLAR | |
COAG | 480 Khz sinusoid with lower output voltage, 80 W @ 100 E |
CUT | 480 Khz sinusoid with high output voltage, 150 W @ 100 E |
Patient Electrode monitoring System (Resistance range) | Dual area PREM patient return Electrode: 10 – 135 ohms |
Single Patient Return electrode: < 20 ohms | |
Power Deviation | less than 15% or 5 watts, whichever is greater |
Dimensions(L X W X H) | 400 x 330 x 120 mm |
Weight | < 7 kg |
Power supply | 180 – 270 Vac , 50/60 Hz |
Max. Power Input | 1000VA, Fuse : 5A |
Standard Accessories: | Optional Accessories: | |
• Monopolar handle with silicon cable with electrode set • Bipolar forcep with silicon bipolar cable • Silicon patient plate with cable • Double Pedal Footswitch • Disposable Hand switch • Mains cord • Unit Cover |
• Reusable Hand switch • Bionet Bipolar forcep • Long 9” Bipolar forcep • Small 4” Bipolar forcep • Angled tip Bipolar forcep • Monopolar forcep with cable |
Intelligent, Smart and Clinical
GEM series of advanced powerful microcontroller based surgical generators gives the surgeon a powerful tool to deliver selected power over varying tissue types. Hence, no need of high power settings and more energy which in turn reduces tissue damage.
- Less thermal damage
- Less charring and sparking
- Reduces recovery time
Micro-controller design
GEM series of generators use a single advanced and powerful controller which controls, senses, and acts upon feedback by continuously monitoring voltage, current, tissue impedance, RF leakage, Power.
- Single advanced and powerful controller.
- Power micro-controller with minimum reaction time and maximum processing speed
- High power efficiency
- Minimizes tissue damage
- Enhanced performance at low power settings which in turn minimizes neuro-muscular stimulation
Surgical Assistance
- User – settable and pre-programmed programs to assist in easy and fast use of the system
- Select the desired program using the program key and the system is ready for use with only one key touch
- Separate program display to avoid confusion with power and modes
Error Indicator
- Performs a self test during switch ON to check system status and resumes in the last used program and modes
- Displays error messages on the display depending upon the faults diagnosed
- Example: Displays error message on display to indicate patient return electrode error” P ERROR”
PPREM
Pad-Patient Return Electrode Monitoring eliminates the risk of patient burns at return electrode site. Pad-patient impedance monitoring using a dual area patient return electrode.
- Audio visual alarm
- Error message on display
- Deactivates the RF to ensure safe operation
Endo cut (Pulse cut)
Endo cut is cutting power delivered in bursts (in seconds) useful in Endo surgeries like papillotomy and polypectomy.
- Three different burst duration in seconds (Two, Three and Four seconds)
Bipolar Sealing
- Optimized combination of pressure and energy creates the seal by melting the collagen & elastin in the vessel wall & reforming it in to a seal & results in no sticking or charring
- Feed back controlled response system automatically discontinues energy delivery when the seal cycle is complete
- High strength seal, Minimal collateral RF thermal effect to adjacent tissue
Modes | |
CUT | Provides uniform cutting through all types of tissues. |
LOW | Provides precise cutting with lower max. voltage which ensures minimum collateral tissue damage suitable for Neuro, Plastic, ENT. |
PURE CUT | Provides pure, precise cutting with higher max. voltage suitable for TURP |
BLEND | Provides CUT power with useful coagulation (cutting with different levels of haemostasis ) to reduce bleeding |
COAG | Satisfies all coagulation needs of the surgeons |
SPRAY | Non contact coagulation with minimal depth of necrosis |
FULGURATE 1 | Provides coagulation with lower max. voltage |
FULGURATE 2 | Provides essential coagulation with high max. voltage with high crest factor |
DESICCATE | Provides low voltage contact coagulation suitable for delicate work |
BIPOLAR | Provides precise bipolar coagulation effect |
MICRO | Precise performance with minimum collateral tissue damage |
MACRO | Provides precise performance with higher maximum voltage |
STANDARD | Provides precise performance, suitable for most applications |
BI CUT | Provides Cutting effect in bipolar using suitable instrument |
BI SEAL | Provides intelligent sealing capabilities to the user upto 7 mm using Open and Lapro attachments, Audio and output cutoff and upon Seal |
SSEG 302
Fully Solid State design
Independent Monopolar isolated outputs
The monopolar generator provides CUT, BLEND1, BLEND2 and SPRAY, FULGURATE, DESICCATE COAGULATION facilities
Operating mode are identified by audio tones and Panel light
Illuminated ON/OFF switch
Patient plate monitoring circuit
Natural convection cooling with silent operation
Bipolar generator with 3 modes
Applications: |
• For all types of minor/major surgeries including under water cutting T.U.R. and for Bipolar micro coagulation. |
Technical Specifications |
|
Monopolar CUT | |
PURE CUT | 400 ± 25 W, 500 ohms load |
BLEND1 | 250 ± 15 W, 500 ohms load |
BLEND2 | 200 ± 15 W, 500 ohms load |
Monopolar COAG | |
SPRAY | 120 ± 10 W, 300 ohms load |
FULGURATE | 150 ± 15 W, 300 ohms load |
DESICCATE | 80 ± 10 W, 300 ohms load |
BIPOLAR | |
MICRO | 80 ± 10 W, 125 ohms load |
MACRO | 80 ± 10W, 125 ohms load |
CUT | 100 ± 10 W, 300 ohms load |
Output Frequency | Frequency 500 ± 25 Khz |
Power | 220-240 Vac ± 5%, 50/60 Hz |
Dimensions(L X W X H) | 400 x 310 x 140 mm |
Weight | < 8kg |
Standard Accessories: | Optional Accessories: | |
• Monopolar handle with silicon cable with electrode set • Bipolar forcep with silicon bipolar cable • Patient Plate with cable • Double Pedal Footswitch • Mains cord • Unit Cover |
• Monopolar forceps with cord and jack • S.S. Paediatric Patient plate with cable • Silicon patient plate with cable • Servo controlled voltage stabilizer • Pencil hand switch ( Disposable / Reusable ) |
SSEG 300B
Fully Solid State design
Independent Monopolar isolated outputs
The monopolar generator provides CUT, BLEND, and COAGULATION facilities.
Operating mode are identified by audio tones and Panel light
Illuminated ON/OFF switch
Patient plate monitoring circuit
Audio-visual indication for patient plate lead faults
Natural convection cooling with silent operation
Bipolar generator has footswitch control
Applications: |
• For all types of minor/major general surgeries and Bipolar micro coagulation. |
Technical Specifications |
|
Monopolar CUT | |
PURE CUT | 300 ± 25 W, 500 ohms load |
BLEND 1 | 200 ± 15 W, 500 ohms load |
BLEND 2 | 200 ± 15 W, 500 ohms load |
Monopolar COAG | |
SPRAY | 120 ± 15 W, 300 ohms load |
FULGURATE | 150 ± 15 W, 300 ohms load |
DESICCATE | 80 ± 15 W, 300 ohms load |
BIPOLAR | |
MICRO | 80 ± 10W, 125 ohms load |
MACRO | 80 ± 10W, 125 ohms load |
CUT | 100 ± 10W, 300 ohms load |
Output Frequency | Frequency 500 ± 25 Khz |
Power | 220-240 Vac ± 5%, 50/60 Hz |
Dimensions(L X W X H) | 400 x 310 x 140 mm |
Weight | <8 kgs |
Standard Accessories: | Optional Accessories: | |
• Monopolar handle with silicon cable with electrode set • Bipolar forcep with silicon bipolar cable • Patient Plate with cable • Double Pedal Footswitch • Mains cord • Unit Cover |
• Monopolar forceps with cord and jack • S.S. Paediatric Patient plate with cable • Silicon patient plate with cable • Servo controlled voltage stabilizer • Pencil hand switch ( Disposable / Reusable ) |
SSEG 402
Fully Solid State design
The monopolar generator provides CUT and COAGULATION facilities
Operating mode are identified by audio tones
Illuminated ON/OFF switch
Patient plate monitoring circuit
Audio indication for patient plate lead faults
Fan cooling with silent operation
Bipolar has CUT as well COAG
Applications: |
• For all types of general surgeries and Bipolar cutting and coagulation. |
Technical Specifications |
|
Monopolar CUT | |
PURE CUT | 200 ± 25 W, 500 ohms load |
Monopolar COAG | |
COAG | 120 ± 15 W, 300 ohms load |
BIPOLAR | |
COAG | 80 ± 10W, 125 ohms load |
CUT | 80 ± 10W, 125 ohms load |
Output Frequency | Frequency 500 ± 25 Khz |
Power | 220-240 Vac ± 5%, 50/60 Hz |
Dimensions(L X W X H) | 270 x 245 x 130 mm |
Weight | 7 kg |
Standard Accessories: | Optional Accessories: | |
• Monopolar handle with silicon cable with electrode set • Bipolar forcep with silicon bipolar cable • Patient Plate with cable • Double Pedal Footswitch • Mains cord • Unit Cover |
• Monopolar forceps with cord and jack • S.S. Paediatric Patient plate with cable • Silicon patient plate with cable • Servo controlled voltage stabilizer • Pencil hand switch ( Disposable / Reusable ) |
SSEG 502
Fully Solid State design
Generator provides CUT and COAGULATION facilities
Operating mode are identified by audio-visual tones
Illuminated ON/OFF switch
Auto Bipolar facility
Natural convection cooling with silent operation
Applications: |
• For all types of general surgeries and Bipolar cutting and coagulation. |
Technical Specifications |
|
BIPOLAR CUT | |
PURE CUT | 80 ± 10 W, 125 ohms load |
BIPOLAR COAG | |
COAG | 80 ± 10 W, 125 ohms load |
AUTO | 80 ± 10 W, 125 ohms load |
Output Frequency | Frequency 500 ± 25 Khz |
Power | 220-240 Vac ± 5%, 50/60 Hz |
Dimensions(L X W X H) | 260 x 210 x 120 mm |
Weight | 5 kg |
Standard Accessories: |
• Bipolar forcep with silicon bipolar cable • Double Pedal Footswitch • Mains cord • Unit Cover |
The GALVOTOM is a modern solid-state electrosurgical unit. Both in design and equipment it meets all the requirements of electrosurgery. High Power and various modes make it ideal for general & clinical practice.
The GALVOTOM SSEG has separate controls for cutting, coagulation, bipolar microcoagulation. It offers all the advantages of electrosurgery: simplified haemostasis (Blend control) with hand as well as foot control.
Its output makes it suitable for procedures in E.N.T., Gynecology, Neurosurgery, Orthopedic, General surgery, Plastic surgeries, Laproscopic Interventions, Including Urological and Thoracic procedures. Can be used for all major and minor operations. Its compact size and light weight makes it a best option for various fields.
RADIO FREQUENCY UNIT
The Technique of Radio Surgery involves the passage of high frequency radio waves(2Mhz) through soft tissue to cut, coagulate, or remove the tissue. Soft tissue resistance to these radio waves causes the cellular water in the soft tissue to heat, which produces steam, an results in cellular molecular dissolution of individual tissue cells.
High frequency radio surgery unit consists of monopolar cutting, Coagulation and Bipolar modes to treat wide variety of skin disorders with excellent cosmetic effects.
Minimum blood loss and pain, Relaxed and easy to use, Minimum heat without charring are some of the features of the Radio Frequency unit.
Technical Specifications |
Input | 220-240 vac, 50/60 Hz | ||||
Frequency | 2 Mhz | ||||
Outputs |
|
||||
Weight | < 3.5 kgs | ||||
Dimensions | 258 x 190 x 82 mm |
Standard Accessories: |
• Monopolar handle with silicon autoclaveable cable • Electrode set • SS patient plate with cable • Single Pedal Footswitch • Unit cover • Mains cord |
RADIO FREQUENCY UNIT RF-2HF
The Technique of Radio Surgery involves the passage of high frequency radio waves(2Mhz) through soft tissue to cut, coagulate, or remove the tissue. Soft tissue resistance to these radio waves causes the cellular water in the soft tissue to heat, which produces steam, an results in cellular molecular dissolution of individual tissue cells.
High frequency radio surgery unit consists of monopolar cutting, Coagulation and Bipolar modes to treat wide variety of skin disorders with excellent cosmetic effects.
Minimum blood loss and pain, Relaxed and easy to use, Minimum heat without charring are some of the features of the Radio Frequency unit.
Works on Handswitch as well as Footswitch
Technical Specifications |
Input | 220-240 vac, 50/60 Hz | ||||
Frequency | 2 Mhz | ||||
Outputs |
|
||||
Weight | < 3.5 kgs | ||||
Dimensions | 258 x 190 x 82 mm |
Standard Accessories: |
• Active handle with Silicon cable • Electrode set • Patient plate with cable • Bipolar Silicon cable with Forcep • Disposable Handswitch • Power cord • Carrying case • Operating manual |
SSEG 302
Made with high quality hypodermic needles – ensures smooth flow of cryogenic gas and last longerEquipped with specially designed 'Silencer' – Enables quieter operation
Body made of ABS plastic, in attractive colors –aesthetically appealing
Specially designed probes for different specific end uses- enables wide usage of same instrument for different specialties
Attractive light-weight carrying case–easy portability
Hand Operated
Uses Nitrous Oxide (N2O) or Carbon dioxide (CO2)
Interchangeable probe facility available
A combined unit for Gynaecology, General Surgery, E.N.T and Dermatological applications with any one probe attached to it
Spare Washers and carrying case
Standard unit is supplied with 1 probe
Probes Available: |
NERVE STIMULATOR (MINI)
Compact
Light Weight
Easy to Use
Nerve stimulator incorporates in a single unit the necessary electronic circuitry for allowing the user to selectively generate twitch pulses at a predetermined low frequency, tetanizing pulses at a relatively higher frequency. Controls are also provided in the unit for adjusting the amplitude of the pulses. The unit further includes suitable indicator means for informing the user as to the particular pulse frequency or pattern being emitted at any given time.
Technical Specifications |
Input Voltage | 230V, 50 Hz |
Output | 0.1 m sec, 1 m sec |
Output Current | 0-5 ma |
Weight | Approx less than 1 Kg |
Dimensions(L X W X H) | 160 x 80 x 50 mm |
Standard Accessories: |
• Mains cable • One set of electrodes with cable • A pair of Velcro straps • Handle with cable with two needle electrodes |
NERVE STIMULATOR (LCD)
Compact, Light Weight, Easy to Use, Aesthetic Look
MICRO-CONTROLLER BASED
MEMBRANE KEY BOARD
ABS SHOCKPROOF BODY
2 LINE LCD DISPLAY
Nerve stimulator incorporates in a single unit the necessary electronic circuitry for allowing the user to selectively generate twitch pulses at a predetermined low frequency, tetanizing pulses at a relatively higher frequency. Controls are also provided in the unit for adjusting the amplitude of the pulses. The unit further includes suitable indicator means for informing the user as to the particular pulse frequency or pattern being emitted at any given time.
Technical Specifications |
Intensity | 0 – 5 ma, Adjustable |
Display | 2 Line LCD Display |
Current Mode | Interrupted |
Pulse | Frequency : 1, 2, 5 Hz, 21 Programs |
Electrodes | 1 P, Handle with cable with electrode – 1 no. |
Resolution | 0.1 ma |
Input | 230 VAC ( Mains Operated ) |
Carrying Case | 1 No. |
ULTRASOUND THERAPY UNIT
(MODEL: ULT10)
Micro-controller based, Feather Touch, Digital System
Uses of ultrasound include facial and body skin rejuvenating treatments, reduction of stretch marks, treatment of contracture and scar tissue such as around breast implants, and the pre- and post-operative treatment of plastic surgery patients to accelerate healing and recovery after procedures such as face lifts, tummy tuck, and liposuction.
Built in 2 line LCD module with backlight
Menu includes continuous mode and pulse mode with
3:4, 2:3, 1:2, 1:4, 1:8, 1:12
42 programs
Computerized panel design with soft touch switches for comfort in operation
User friendly, simple and easy to operate
Built in clinical programs
Technical Specifications |
Display | 16 X 2 LCD |
Output Timer | 2.5 W/cm2 |
Input Supply | 230 VAC, 50 Hz |
Modes | Continuous, Pulse Modes |
Frequency | 1Mhz |
Pulse Frequency | 8, 48, 100Hz |
Duty Cycle | 10, 20, 50% |
Timer | 0 – 30 MIN |
Dimension | 275 x 200 x 85 mm |
Standard Accessories: |
• Light Weight Sound Head with Cable • Supplied with Carrying Case and gel bottle |
HALOGEN SPOT LIGHT
MODEL : 12V, 50W
Useful in Surgery, Gynaecology, ENT and Dermatology
Variable regulator to control intensity
Pedestal (Floor) as well as wall mounted models
Light intensity of 1250 foot candles at 16 in. diameter
Light temp. is kept at 3400 Kelvin to enable identification of true
tissue color
Adjustable in Any direction, angle and height
Extremely Compact and light weight
Source gun is Detachable
Easy to Dismantle and Reassemble
Ideal for Labour room, OPD and Examination
Suitable for Examination in Labour Room/OPD or in Operation theatre if additional light is required. Single 15w Power LED Module, hence low power consumption. Sleek Design. Also available Wall Mount Clamp (optional) for wall mounting the light.
Light Intensity is adjustable
White light, No Heat
Flexible Gooseneck for Positioning
Life more than 50000hrs
Input Supply : 180 – 270Vac
Mobile on Wheels, Sleek Design
SS stand, Height Adjustable
Lux more than 50000
Colour temp 5500k
Wet-Field Bipolar Coagulator is a mini diathermy solid state for controlling bleeding during surgical procedures, eliminating tissues trauma using Bipolar standard Forceps and Erasers.
Tissue may be coagulated under irrigation (wet-field) with saline or in a bloody field. The high frequency current passes from one forcep tip to the other coagulating only the bleeder or particular tissue.
Important Features
- Foot controlled system gives convenience to surgeons.
- Forceps, Erasers and Cables are autoclavable.
- Digital display for easy setting.
- Imported Forcep and Erasers can be attached.
- Audible tone lets surgeon know the unit is activated.
- Compact in size, elegant look and portable.
Benefits
- Solid state circuit system ensures safest procedures to delicate tissue involved for Cataract, Retinal and IOL Surgery.
- Can be used with irrigation (wet-field) or in the bloody field itself, to locate the bleeder easier to enable the surgeon to have better visualisation.
- Useful for Ophthalmic surgical procedures and for delicate surgery reducing tissue trauma.
- Coagulation with saline irrigation reduces tissue heating, shrinkage, drying and sticking of coagulam to the forcep tips.