INFANT WARMER WITH DOUBLE SURFACE PHOTOTHERAPY

UPPER SURFACE PHOTOTHERAPY STAND
Sturdy mobile stand with height adjustment and easy swivelling of source box
Unit can be used along with Infant Care Trolley, Radiant Warmer and Incubator.

  • Four, fluorescent lamps (CFL) emitting blue radiation with a peak at 450 nm specially used for hyperbilirubinaemia treatment
  • Two day-light (CFL) tubes for observation
  • Protective grill
  • Mirror coated reflector
  • Irradiance approx. 12-15 µwatts/cm/nm 2
  • Source Cooling Fan
  • Optional Time Totaliser

INFANT TROLLEY

Sturdy mobile crib mounted on castors can be used along with warmer stand, wall mounted warmer, phototherapy stand & as a separate bed for neonates

  • Easy access of an Infant from all sides due to clear acrylic collapsible side panels
  • Baby tray with transparent slot to incorporate the facility of Under Surface Phototherapy
  • Acrylic Baby Tray for infant’s head Up / Down facility
  • Provision for positioning X-Ray cassette
  • Two Lower shelf
  • Thick cushion mattress
  • Movable on castors
     







UNDER SURFACE PHOTOTHERAPY

Detachable source box can be used  along with radiant warmer and infant crib

  • Four, fluorescent lamps (CFL) emitting blue radiation with a peak at 450 nm specially used for hyperbilirubinaemia treatment
  • Two day-light (CFL) tubes for observation
  • Protective grill
  • Mirror coated reflector
  • Irradiance approx. 12-15 µwatts/cm /nm
  • Source Cooling Fan
  • Optional Time Totaliser
 

  Skin / Manual

    Skin / Air / Manual
  • 1” bright skin temp. display
  • Set temp. display
  • % of heater output display on Bar Graph
  • Selection of Skin & Manual modes
  • Temperature display in o C or o F
  • APGAR Timer
  • Feed timer
  • Audio Alarm mute facility
  • Mute time programmable

Manual Mode

  • Time programmable
  • Heater output adjustable
  • Over temp. cutoff.

 Skin Mode
Safety alarms
(with status LED)
Temp. High (>1°C of set temp.)
Temp. Low (<1°C of set temp.)
Temp. Over Range (>38°C)
Temp. Under Range (<34°C)
(Actual temp. starts blinking)
Probe Fail
Power failure
Heater Failure
System Failure

Safety high temp. cutoff at 38°C
  • 1” bright skin temp. display
  • 1” bright air temp. display
  • % of heater output display on Bar Graph
  • Separate Set temp. display
  • Temperature display in o C or o F
  • APGAR Timer
  • Feed Timer
  • Baby attending alarm
  • Mute time programmable
  • Audio Alarm mute facility

Manual Mode

  • Time programmable
  • Heater output adjustable
  • Over temp. cutoff.

Safety alarms (with status LED)
(Actual temp. starts blinking)

Skin Mode
Skin Temp. High (>1.5°C of set temp.)
Skin Temp. Low (<1.5°C of set temp.)
Skin Temp. Over Range (>38°C)
Skin Temp. Under Range (<34°C)
Skin Probe Fail

Air Mode
Air Temp. High (>1.5°C of set temp.)
Air Temp. Low (<3.0°C of set temp.)
Air Temp. Over Range (>39°C)
Air Probe Fail
 
Safety high temp. cutoff at

  • Skin Temp.> 38oC
  • Air Temp.> 39oC
  • Power Fail
  • Heater Fail
  • System Fail